(item no. 4459442)
Product Overview
The Fluke a3002 FC Wireless AC/DC Current Module is one member of a family of wireless test tools that let you share test data using the Fluke Connect app with ShareLive video call. Fluke Connect Wireless Test Tools put tools, not your body, near live electrical panels. It’s simple. Connect the a3002 FC current module, or any of the other Fluke Connect remote modules, and read the results on the Fluke Connect Wireless Multimeter or a portable PC from a safe distance.
The a3002 FC AC/DC current module wirelessly relays measurements to other Fluke Connect® enabled master units, listed below:
- Fluke 3000 FC DMM
- Fluke Ti200/300/400Thermal Camera
- PC via optional pc3000 FC Adapter
- Fluke Connect Mobile app
Features
The Fluke a3002 FC Wireless AC/DC Current Clamp Module is designed to be used with the i410 or i1010 AC/DC current clamps (not included with module). The a3002 FC AC/DC current module wirelessly relays measurements to other Fluke Connect® enabled master units, listed below:
- Fluke 3000 FC DMM
- Fluke Ti200/300/400Thermal Camera
- PC via optional pc3000 FC Adapter
- Fluke Connect Mobile app
The Fluke a3002 FC main features include:
- Measure up to 400 AAC or 400 ADC with i410 (sold separately)
- Measure up to 600 AAC or 1000 ADC with i1010 (sold separately)
- Use as a standalone meter or as part of the system
- Logging function for recording and saving up to 65,000 readings
Specifications: Fluke a3002 FC Wireless AC/DC Current Module
AC/DC Current Function | ||
A DC | Range / resolution | 1000 A / 0.1 A |
Accuracy1 | (0.5% + 3) | |
A AC | Range / resolution | 600.0 A / 0.1 A |
Accuracy1 | (1.0% + 3) | |
1. AC accuracy is specified from 45 Hz to 1000 Hz | ||
Specifications | ||
Maximum voltage between terminal and earth ground (module) | 30 V DC or AC RMS | |
LCD w/backlight | 3½ digits, 6000 counts, 4/sec update rate | |
Log rate/Interval | Adjustable by PC from 1 sec to 1 hr, default 1 min | |
Battery type | 2 AA Alkaline batteries, NEDA 15A, IEC LR6 | |
Battery life | 400 hours minimum | |
Memory | Record up to 65,000 readings | |
RF communications | 2.4 GHz ISM Band | |
RF communication range | Open air, unobstructed | Up to 20 m |
Obstructed, sheetrock wall | Up to 6.5 m | |
Obstructed, concrete wall or steel electrical enclosure | Up to 3.5 m | |
Operating temperature | -10°C to +50°C | |
Storage temperature | -40°C to +60°C | |
Temperature coefficient | 0.1 X (specified accuracy) / °C (< 18°C or > 28°C) | |
Relative humidity | 0% to 90% (0°C to 35°C) 0% to 75% (35°C to 40°C) 0% to 45% (40°C to 50°C) |
|
EMC | IEC 61236-1, Portable | |
Safety compliance | Protected to IEC 61010-1, 600 V CAT IV / 1000 V CAT III, 3rd edition Pollution Degree 2 | |
Certifications | CSA, CE, FCC: T68-FBLE IC: 6627A-FBLE | |
Ingress Protection (IP) rating | IP42 | |
Size (HxWxD) | 160 x 66 x 38 mm (6.3 x 2.6 x 1.5 in) | |
Weight | 0.255 kg (9 oz) | |
Warranty | Three years | |
Not compatible with Fluke CNX test tools |
Includes:
- True-RMS AC/DC Current Module
- Magnetic hanging strap
Manuals
- v3000 FC | Calibration manual
- v3000 FC v3001 FC | Safety sheet
- v3000 FC v3001 FC | Quick reference guide
Data Sheets
- Fluke 3000 FC DMM Data Sheet